Steam Sterilization:
After cleaning metal part of GSV200.DBL. Each metal part GSV200.DBL is load in separate self-sealed pouches:
| Steaming types |
Gravity Displacement |
Pre-Vacuum |
| Enzymatic Cleaning Solution like prolystica |
Yes |
Yes |
| Temperature |
132 C |
132 C |
| Cycle Time (minutes) |
15 |
4 |
| Dry time (minutes) |
20-30 |
20-30 |
| Method |
Camera arrangement |
Metal part laryngoscope unit |
| Enzymatic Cleaning Solution like prolystica |
Yes |
Yes |
| Neutral pH soap water |
Yes |
Yes |
| CidexOPA |
Yes |
No |
| Steris Process |
Yes |
No |
| Autoclave, steam or heat |
No |
Yes |